Page 24 SMV800 Modbus User's Guide Revision 1.0
See Table 2-3 and Table
2-4 for Mass Flow and
Volume Flow units
Allows configuring Flow unit.
All the units are self-Explanatory.
Custom Unit:
When this unit is selected, Tools will populate
Flow Custom Tag
Flow Base Unit : Base unit is unit from which
custom unit is derived
Flow Conver. Factor: Enter a numeric value
that represents the number of base units per
one
custom unit.
Example:
Flow Custom Tag: MyNewUnit
Flow Base Unit: g/sec
Flow Conver Factor: 0.5 (means 0.5 g/sec = 1
Custom Unit)
Flow Rate = 50 g/sec
Flow Rate in “MyNewUnit” will be = (50/0.5)
MyNewUnit
Write a new Lower Range Value and Upper
Range Value for the Flow input.
Flow can be ranged anywhere between –URL
to +URL so that the span is <= URL (i.e.,URV
– LRV should be <= URL)
The lower value for Low Flow cutoff. When the
flow drops below this value, the flow output will
be forced to 0%.
The upper value for Low Flow cutoff. The flow
will not exit the low flow cutoff state (0% flow)
until the flow exceeds this value.
Flow Cutoff Lo Lmt
Flow Cutoff Hi Lmt
Allows the user to configure new values for the
low and high cutoff limits for the Low Flow
Cutoff option