EasyManua.ls Logo

Honeywell SMV800 - Table 2-6: Sensor Types for Process Temperature Input

Honeywell SMV800
220 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Page 32 SMV800 Modbus User's Guide Revision 1.0
Sensor Type
Identify and select the type of sensor that is connected to the transmitter as its input for process
temperature. This will set the appropriate LRL and URL data in the transmitter automatically.
Table 2-6 shows the pre-programmed temperature sensor types and the rated measurement range
limits for a given sensor selection.
Table 2-6: Sensor Types for Process Temperature Input
Input
Type
Maximum Range Limits
RTD
(2,3,4 wire)
° C
° F
Pt254
-200 to 850
-328 to 1562
Pt100
-200 to 850
-328 to 1562
Pt2004
-200 to 850
-328 to 1562
Pt5004
-200 to 850
-328 to 1562
Pt10004
-200 to 500
-328 to 932
Thermocouples
° C
° F
B4
200 to 1820
392 to 3308
E
-200 to 1000
-328 to 1832
J
-200 to 1200
-328 to 2192
K
-200 to 1370
-328 to 2498
N4
-200 to 1300
-328 to 2372
R4
-50 to 1760
-58 to 3200
S4
-50 to 1760
-58 to 3200
T
-250 to 400
-418 to 752
ATTENTION: Whenever you connect a different sensor as the
transmitter’s input, you must also change the sensor type
configuration to agree. Otherwise, range setting errors may result.