Page 126 SMV800 Modbus User's Guide Revision 1.0
ps_no_factory_c
alibration
Factory Calibration for the
Pressure
module is missing.
Accuracy will be
compromised.
Return the device for
Factory Calibration
Internal communication
quality between the
Communication Module
and Pressure Sensor is
degrading.
Either the transmitter is
installed in a noisy
environment or internal
communication quality
between the
Communication Module
and Pressure Sensor
module is degrading.
Verify the connector for
bent pins.If
cable/connector is not the
problem, replace the
Communication module.If
problem still persists,
replace the Terminal
board.
ps_sensor_over
_temperature
The Meter Body
temperature is too high.
Accuracy and life span
may decrease if it
remains high.
Verify the environmental
temperature is within
specification. Take steps to
insulate the
Temperature module from
the temperature
source.
NVM copy of Common
database block found
corrupt.
Power cycle the device. If
the problem persists,
replace the transmitter
NVM copy of Vital
Configuration database
block found corrupt.
Power cycle the device. If
the problem persists,
replace the transmitter
general_config_
db_corrupt:1;
NVM copy of General
Configuration database
block found corrupt.
Power cycle the device. If
the problem persists,
replace the transmitter
NVM copy of Configuration
Change database block
found corrupt.
Power cycle the device. If
the problem persists,
replace the transmitter
NVM copy of Advanced
Configuration database
block found corrupt.
Power cycle the device. If
the problem persists,
replace the transmitter
NVM copy of Display
View/Screens
Configuration database
block found corrupt.
Power cycle the device. If
the problem persists
reconfigure Display Screen
parameters. If still problem
is seen replace
communication module.