Page 146 SMV800 Modbus User's Guide Revision 1.0
Compensation
Switch
Shows currently selected
Compensation
settings for Static Pressure
and Temperature..This is
user configurable parameter.
(Byte 0 :-> Bit 6 to 7 )
Bit 6 :-> 0 = Off, 1 = AP
Compensation On
Bit 7 :-> 0 = Off, 1 = Temp
compensation On
Primary
Element Type
288
Shows currently selected
Primary Element Type.This
is user configurable
parameter. Available Primary
Element options are
dependent upon the selected
Algorithm Option; Advanced
Algorithms or ASME 1989
Algorithms.
R/W
uint8
Primary Element Type
For Legacy Control =
SMV3000 Method:
0x00 = O-FTaps(ASME-
ISO)D >/= 2.3
0x01 = FTaps(ASME-
ISO)2</= D </= 2.3
0x02 = O-CTaps(ASME-
ISO)
0x03 = O-
D&D/2Taps(ASME-ISO)
0x04 = O-
2.5D&8DTaps(ASME-ISO)
0x05 = V-MI(ASME-ISO)
0x06 = V-RCI(ASME-ISO)
0x07 = V-RW Sheet-Iron
Inlet(ASME-ISO)
0x08 = N-(ASME LR)
0x09 = V-nozzle(ISA Inlet)
0x0A = V-Leopold
0x0B = V-Gerand
0x0C = Uni V-Tube
0x0D = Lo-Loss V-Tube
0x0E = PE-0.875inch for
2inch Pipe
0x0F = PE-0.875inch for
2.5inch Pipe
0x10 = PE-0.875inch for
3inch Pipe
0x11 = PE-0.875inch for
4inch Pipe
0x12 = PE-0.875inch for
5inch Pipe
0x13 = PE-0.875inch for
6inch Pipe
0x14 = PE-0.875inch for
8inch Pipe
0x15 = PE-0.875inch for
10inch Pipe
0x16 = PE-0.875inch for
12inch Pipe
0x17 = PE-0.875inch for
14inch Pipe
0x18 = PE-1.25inch for
12inch Pipe
0x19 = PE-1.25inch for
14inch Pipe
0x1A = PE-1.25inch for
16inch Pipe
0x1B = PE-1.25inch for
18inch Pipe
0x1C = PE-1.25inch for
20inch Pipe
0x1D = PE-1.25inch for
22inch Pipe
0x1E = PE-1.25inch for

Table of Contents