Page 166 SMV800 Modbus User's Guide Revision 1.0
Static
Pressure URV
during factory
B calibration
809
The SP URV trim point at
which factory B calibration
was performed.
RO
Float
N/A
Static
Pressure LRV
during factory
B calibration
811
The SP LRV trim point at
which factory B calibration
was performed.
RO
Float
N/A
Static
Pressure URV
during factory
C calibration
813
The SP URV trim point at
which factory C calibration
was performed.
RO
Float
N/A
Static
Pressure LRV
during factory
C calibration
815
The SP LRV trim point at
which factory C calibration
was performed.
RO
Float
N/A
RESERVE_F
OR_OID_CHA
NGE_2HR
817
NOT USED
RO
DP max
819
Maximum value of DP
recorded with a time and
date stamp with units of
seconds since the event
RO
Float
N/A
DP min (not
used in AP or
GP devices)
821
Minimum value of DP
recorded with a time and
date stamp with units of
seconds since the event
RO
Float
N/A
SP max (not
used in AP or
GP)
823
maximum value of SP
recorded with a time and
date stamp with units of
seconds since the event
RO
Float
N/A
MB Temp max
825
Maximum value of meter
body temperature recorded
with a time and date stamp
with units of seconds since
the event
RO
Float
MB Temp min
827
Minimum value of meter
body temperature recorded
with a time and date stamp
with units of seconds since
the event
RO
Float
Pressure
Sensor
Controller
Core Temp
max
829
Maximum value of
temperature recorded for the
controller on pressure meter
body with a time and date
stamp with units of seconds
since the event
RO
Float
Pressure
sensor
controller Core
Temp min
831
Minimum value of
temperature recorded for the
controller on pressure meter
body with a time and date
stamp with units of seconds
since the event
RO
Float