38 Chapter1
General Information
Recommended Test Equipment
Connectors: BNC(f)
20 dB fixed Frequency range: dc to 18 GHz HP 8491B
P,V
attenuator Attenuation accuracy: <±1 dB Option 020
Maximum SWR: 1.2 (dc to 2.9 GHz)
10 dB fixed Frequency range: dc to 18 GHz HP 8491B
P,V
attenuator Attenuation accuracy: <±0.6 dB Option 010
Maximum SWR: 1.2 (dc to 2.9 GHz)
Reference
attenuator
supplied with HP 8481D HP 11708A
P,A
Termination Frequency range: dc to 2.9 GHz HP 908A
P,M,V
Impedance: 50 Ω
Maximum SWR: <1.10
Connector: Type N(m)
Low-pass filter Cutoff frequency: 50 MHz 0955-0306
P,M,V
Rejection at 65 MHz: >40 dB
Rejection at 75 MHz: >60 dB
Power splitter Frequency range: 1 kHz to 12 GHz HP 11667A
P,A,M,V
Insertion loss: 6 dB (nominal)
Output tracking: <0.25 dB
Equivalent output SWR: <1.22
Service
accessory kit
No substitute 08562-60021
A,T
Tuning tool No substitute 8710-1010
A
Cables
Test cable Connectors: BNC (m)-to-SMB (f) 85680-60093
A,M
Length: ≥61 cm (24 in.)
Cable,
RG-214/U
Connectors: Type N (m) HP 11500A
P,V
Length: ≥91 cm (36 in.)
Cable Connectors: SMA (m) 8120-1578
P
Length: 24 to 36 inches
* Part of microwave workstation
P = performance tests; A = adjustments; M = test & adjustment module; T = troubleshooting;
V = operation verification
Table 1-3 Recommended Test Equipment
Instrument Critical Specifications for Equipment Substitution Recommended
Model
Use