EasyManua.ls Logo

HP 8712ES - Page 436

HP 8712ES
507 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
9-26 ES User’s Guide
Specifications
Table 9-7 Test Port Output (continued)
a. Maximum power setting is the highest power setting allowed that maintains the network
analyzer’s specified level accuracy.
b. Option 1E1 adds a 60 dB step attenuator to the HP 8712ET/8714ET.
HP 8712ET/ES and HP 8714ET/ES Test Port Output
Description Specification
Supplemental
Information
Output Power (continued)
<1 GHz >1 GHz
Maximum Power Setting:
a
HP 8712ET:
50 :
With Attenuator Option 1E1
b
15 dBm 12 dBm
Without Attenuator Option 1E1 16 dBm 13 dBm
75 :
With Attenuator Option 1E1
b
12 dBm 9 dBm
Without Attenuator Option 1E1 13 dBm 10 dBm
HP 8712ES:
50 13 dBm 10 dBm
75 10 dBm 7 dBm
HP 8714ET:
50 :
With Attenuator Option 1E1
b
10 dBm 9 dBm
Without Attenuator Option 1E1 11 dBm 10 dBm
75 :
With Attenuator Option 1E1
b
7 dBm 6 dBm at
<2 GHz
6 dBm at >2 GHz,
char.
Without Attenuator Option 1E1 8 dBm 7 dBm at
<2 GHz
7 dBm at >2 GHz,
char.
HP 8714ES:
50 9 dBm 7 dBm
75 6 dBm 4 dBm at
<2 GHz
4 dBm at >2 GHz,
char.

Table of Contents

Other manuals for HP 8712ES

Related product manuals