Chapter 7 7-19
Error Messages
201 to 799: Device-Specific Errors
201 to 799:
Device-Specific Errors
Some device operations did not properly complete, possibly due to an
abnormal hardware or firmware condition. These codes are also used
for self-test response errors. Errors in this class set the device-specific
error bit (bit 3) in the event status register (IEEE 488.2, section 11.5.1).
The <error_message> string for a positive error is not defined by SCPI.
A positive error indicates that the instrument detected an error within
the HP-IB system, within the instrument firmware or hardware, during
the transfer of block data, or during calibration.
653 Auto Align not available when using
Calibration Defaults
The Auto Alignment system cannot be used until an
Align Now All is executed by pressing System,
Alignments, Align Now, All. For HP E4402B, E4403B,
E4404B, E4405B, E4407B, and E4408B only you
must connect the
AMPTD REF OUT to the INPUT with
the appropriate cable to perform this alignment.
614 Bad or missing floppy disk
The floppy is not inserted or the directory could not be
read. Insert a known good disk and try again.
205 Command not recognized
Indicates that the command sent from the remote
interface was not recognized. Check the programming
guide for correct syntax.
219 Command not valid in this model
Indicates that the command sent from the remote
interface does not apply to this model number. For
example, attempting to center the preselector in an
analyzer without a preselector will generate this error.