7-20 Chapter7
Error Messages
201 to 799: Device-Specific Errors
652 Connect Amptd Ref Output to Input
For HP E4402B, E4403B, E4404B, E4405B, E4407B,
and E4408B only: you must connect the AMPTD REF
OUTPUT to the analyzer INPUT with the appropriate
cable.
651 Connect RF OUT to INPUT
Attempt to align the tracking generator without its
output connected. Connect the tracking generator RF
OUT to the analyzer INPUT.
615 Corrupted file
The file that you were trying to load is corrupt.
610 File access is denied
The file is protected or hidden and cannot be accessed.
604 File already exists
Attempt to save to a file that already exists. Delete or
rename the old file and try again.
607 File Name Error
An invalid file name has been specified. Use filenames
with a maximum of 8 characters (letters and digits
only) and use a 3 character extension. Note that
lowercase and uppercase are perceived as the same.
612 File not found
The analyzer could not find the specified file.
613 Flash memory is full
The internal flash memory is full. Clear some space by
deleting unwanted files.You may also increase the flash
memory size by purchasing Option B72.