17
INSULATION RESISTANCE TEST MODE
Voltage Setting
Range:
30 1000V DC
Resolution:
1V
Accuracy:
(2 of setting 5 volts)
Voltage Display
Range:
0 1000 V
Resolution:
1V
Accuracy:
(2 of reading 2 counts)
Resistance Display
Range:
1 9999 M (4 Digit, Auto Ranging)
Resolution:
500VDC - 1000VDC
M
M
0.001
1.000 - 9.999
0.01
10.00 - 99.99
0.1
100.0 - 999.9
1
1000 - 9999
Accuracy:
(2 of reading 2 counts) at test
voltage
500 1000V and 1 999.9 M
(5 of reading 2 counts) at test
voltage
500 1000V and 1000 9999 M
(8 of reading 2 counts) at test
voltage
30 500V and 1 1000 M
Maximum Limit
Range:
0,1 9999 M (0 = Off)
Resolution
1M
Accuracy
Same as Resistance display
Minimum Limit
Range:
1 9999 M
Resolution
1M
Accuracy
Same as Resistance display
Short Circuit Current
Maximum:
9mA-peak
Ramp Timer
Range:
Resolution:
Accuracy:
Ramp-Up: 0.1 999.9 sec
Ramp-Down: 1.0 999.9 sec 0=OFF
0.1 sec
(0.1% of reading + 0.05 sec)
Delay Timer
Range:
0, 0.5 999.9 sec (0 = Constant)
Resolution:
0.1 sec
Accuracy:
(0.1% of reading + 0.05 sec)
Ground Fault Interrupt
GFI Trip Current: 450A max
HV Shut Down Speed: <1ms

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the IKONIX Associated Research Hypot III and is the answer not in the manual?

IKONIX Associated Research Hypot III Specifications

General IconGeneral
BrandIKONIX
ModelAssociated Research Hypot III
CategoryMeasuring Instruments
LanguageEnglish