Figure 12 Remove from Clamshell Package
4 Clean the glass surface of the flow cell with a lint-free alcohol wipe. Dry the glass with a low-lint lab
tissue.
NOTE
Make sure the glass surface of the flow cell is clean. If necessary, repeat the cleaning step.
5 Align the flow cell over the alignment pins and place the flow cell on the stage.
Figure 13 Load the Flow Cell
6 Select Load.
The door closes automatically, the flow cell ID appears on the screen, and the sensors are checked.
NOTE
Keep hands clear of the flow cell door while it is closing to avoid pinching.
7 Select Next.
Empty the Spent Reagents Container
1 Open the buffer compartment door with the latch under the lower-left corner of the door.
2 Remove the spent reagents container and discard the contents in accordance with applicable
standards.
Document # 1000000041922 v01
For Research Use Only. Not for use in diagnostic procedures.