Chapter 5 Maintenance
Introduction 35
Perform a Manual Wash 35
Replace Air Filter 38
Software Updates 39
Reboot and Shut Down Options 40
Introduction
Maintenance procedures include manual instrument washes and system software updates when
available.
u Instrument washesAn automatic post-run wash after each sequencing run maintains instrument
performance. However, a manual wash is required periodically under certain conditions. See
Perform
a Manual Wash
on page 35.
u Software updatesWhen an updated version of the system software is available, you can install
the update automatically through a connection to BaseSpace or manually after downloading the
installer from the Illumina website. See
Software Updates
on page 39.
u Air filter replacementRegular replacement of the air filter ensures proper air flow through the
instrument.
Preventive Maintenance
Illumina recommends that you schedule a preventive maintenance service each year. If you are not
under a service contract, contact your Territory Account Manager or Illumina Technical Support to
arrange for a billable preventive maintenance service.
Perform a Manual Wash
Manual washes are initiated from the Home screen. Wash options include the Quick Wash and the
Manual Post-Run Wash.
Wash Types Description
Quick Wash
Duration: 20 minutes
Flushes the system with a user-supplied wash solution of laboratory-grade water and
Tween 20 (buffer wash cartridge).
Required every 14 days that the instrument is idle with reagent cartridge and buffer
cartridge in place.
Required every 7days that the instrument is in a dry state (reagent cartridge and
buffer cartridge removed).
Required after a shutdown.
Manual Post-Run Wash
Duration: 90 minutes
Flushes the system with a user-supplied wash solution of laboratory-grade water and
Tween 20 (buffer wash cartridge) and 0.12% sodium hypochlorite(reagent wash
cartridge).
Required if the automatic post-run wash was not performed.
A manual wash requires the reagent wash cartridge and buffer wash cartridge provided with the
instrument, and a used flow cell. A used flow cell can be used up to 20 times for instrument washes.
Document # 1000000041922 v01
For Research Use Only. Not for use in diagnostic procedures.
35

Other manuals for illumina NextSeq 550Dx

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the illumina NextSeq 550Dx and is the answer not in the manual?

illumina NextSeq 550Dx Specifications

General IconGeneral
Brandillumina
ModelNextSeq 550Dx
CategoryLaboratory Equipment
LanguageEnglish

Related product manuals