2 Select Next.
3 Confirm run parameters.
u Run NameName of the run as assigned in BaseSpace.
u Library IDName of the pooled libraries as assigned in BaseSpace.
u RecipeName of the recipe, either NextSeq High or NextSeq Mid depending on the reagent
cartridge used for the run.
u Read TypeSingle Read or Paired End.
u Read LengthNumber of cycles for each read.
u [Optional] Custom Primers, if applicable.
4 [Optional] Select the Edit button to change run parameters. When finished, select Save.
u Run parametersChange the number of reads or number of cycles per read.
u Custom primersChange the settings for custom primers. For more information, see
NextSeq
Custom Primers Guide (document # 15057456)
.
u Purge consumables for this runChange the setting to purge consumables automatically after
the current run.
5 Select Next.
Enter Run Parameters (Standalone Configuration)
1 Enter a run name of your preference.
2 [Optional] Enter a library ID of your preference.
3 Select a read type, either Single Read or Paired End.
4 Enter the number of cycles for each read in the sequencing run.
u Read 1Enter a value up to 151 cycles.
u Index 1Enter the number of cycles required for the Index 1 (i7) primer.
u Index 2Enter the number of cycles required for the Index 2 (i5) primer.
u Read 2Enter a value up to 151 cycles. This value is typically the same number of cycles as
Read 1.
The control software confirms your entries using the following criteria:
u Total cycles do not exceed the maximum cycles allowed
u Cycles for Read 1 are greater than the 5 cycles used for template generation
u Index Read cycles do not exceed Read 1 and Read 2 cycles
5 [Optional] If you are using custom primers, select the checkbox for the primers used. For more
information, see
NextSeq Custom Primers Guide (document # 15057456)
.
u Read 1Custom primer for Read 1.
u Index 1Custom primer for Index 1.
u Index 2Custom primer for Index 2.
u Read 2Custom primer for Read 2.
6 [Optional] Select the Advanced Settings button to change run parameters.
u From the Recipe drop-down list, select a recipe. Only compatible recipes are listed.
u Output folder locationChange the output folder location for the current run. Select Browse to
navigate to a network location.
u Included fileSelect files to include in the Output Folder that can be useful if further analysis is
required. For example, manifest files and sample lists.
Document # 1000000041922 v01
For Research Use Only. Not for use in diagnostic procedures.
23
NextSeq 550Dx Research Mode Instrument Reference Guide

Other manuals for illumina NextSeq 550Dx

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the illumina NextSeq 550Dx and is the answer not in the manual?

illumina NextSeq 550Dx Specifications

General IconGeneral
Brandillumina
ModelNextSeq 550Dx
CategoryLaboratory Equipment
LanguageEnglish

Related product manuals