Page #16 background imageLoading...
Page #16 background image
page 15
2.Data Output Formats
Data Output Format [Q][±fffff][±ddddddd][P][L][C][S][X][CR]
Memory Data Output Format m[±fffff][±ddddddd][P][L][C][S][YY][MM][DD][hh][mm][ss][CR]
Q Type of Data f: Continuous data output (2000 data/sec)
I: Continuous data output (10 data/sec)
a: Continuous +peak data
h: Continuous –peak data
r: Real time data
p: +Peak
n: –Peak
1: 1st Peak
2: 2nd Peak
±fffff Force Data w/decimal point
±ddddddd Displacement Data +0000000 (placeholder)
P Selected Unit # 0 - 5
L Displacement Unit 0 (placeholder)
H: +NG
C Comparator O: OK
L: –NG
E: Overload
S 2nd/3rd Setpoints 0 (placeholder)
0: No input
1: No Rec input and Mark input
X Signal Line Conditions 2: Rec input and No Mark input
(Rec Signal and Mark Signal) 3: Rec input and Mark input
4: Rec + Shift input and No Mark input
5: Rec + Shift input and Mark input
YY Year
MM Month
DD Date/Time Stamp Date
hh Hour
mm Minute
ss Second
Q Type of Data
C High/Low Setpoints
X Signal Line Conditions
(Rec Signal and Mark Signal)
Date/Time Stamp
3.Force Unit Codes
FORCE UNIT CODES
CODE 00 01 02 03 04 05 07 08 10 11 12 13 14 16 17 22 23
UNIT None mN N kN g kg gf kgf ozf lbf klbf N-cm N-m kgf-cm kgf-m ozf-in lbf-in

Other manuals for Imada ZTS Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Imada ZTS Series and is the answer not in the manual?

Imada ZTS Series Specifications

General IconGeneral
DisplayLCD
Sampling Rate2000 Hz
Measurement UnitsN, kgf, lbf
Data OutputUSB, RS-232
Power SupplyAC adapter or battery
Battery LifeApprox. 8 hours
Operating Temperature0°C to 40°C
Storage Temperature-10°C to 50°C (14°F to 122°F)
Accuracy+/- 0.2% F.S.