Equipment Settings 67
Gas type
For editing the gas equivalent go to the
GASline item and press “EDIT”. Only gas
equivalents that are listed in the
GAS/TRIGGER submenu (Gas1 to Gas4) are
available for selection. The gas equivalent is stated in plain text in parenthesis. The
list of maximum four available gas equivalent can be scrolled with the up and down
buttons.
Notice The I•Guide Mode refers to the gas equivalent selected in the “
GAS/
TRIGGER” submenu. If you change a gas equivalent in the “GAS” submenu
the gas equivalent selected for the I•Guide program will automatically
change as well.
No. of points
The number of potential leak locations to be checked can be varied between 0 and
99 and can be edited with the up and down keys.
Default value: 4
Notice The I•Guide mode can also be used as a timer signal only. If the number of
points is set to 0, the Protec P3000 will prompt for the next location to be
tested continuously without using the global leak rate function.
Notice The I•Guide mode can also be used to summarize leak rates on demand.
If the number of points is set to 99, a results screen with summarized global
leak rate will be displayed after the right button has been pressed for 2 s
continiously (or after the 98th point automatically).
Measuring time
The
MEASURING TIME during which the sniffer tip needs to be hold to the correct loca-
tion can be programmed to values between 0.7 and 25.0 s. The measuring time
should not be less than the response time of the Protec P3000. We recommend to
use the following minimum measuring times:
Please use the
UP and DOWN buttons for editing this parameter.
Default value: 1.0 s
Wait time
The
WAIT TIME between two leak checks allowed to move the sniffer to the next
potentially leaking location can be varied between 0.1 s and 25.0s in steps of 0.1s .
Please use the
UP and DOWN buttons for editing this parameter.
Default value: 3.0 s
Sniffer line length min. measuring time
3m 0.7 s
5m 0.9 s
10m 1.4 s
15m 3.0 s