EasyManuals Logo

Inficon Protec P3000 User Manual

Inficon Protec P3000
128 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #74 background imageLoading...
Page #74 background image
74 Equipment Settings
Operation hours h
Serial number 11-digit number
Wise Serial number
Software version x.xx.xx
Circuit temperature °C
Temperature of main
board
Test leak temperature °C
Sensor block Protec P3000 1 / 2 / 3
1 = Protec P3000
2 = Protec P3000 with
extended measurement
range
3 = Protec P3000XL
Page 2: Sensor Global
I akt. A
Leak rate mbar l/s helium leak rate
PWM-Membrane Integer
Heater Voltage V
High Voltage V
Operating time Wise h
Wise Temperature °C
Software Version
Wise Sensor
Wise-State Integer
Measure Background Press button
If pressing the
MEASURE BACKGROUND button the
Protec P3000 will go into a special measurement
mode and determine the background concentration
of Helium currently prevalent in the environment.
This process may take a couple of seconds. This
feature is not a continuous signal display but should
be used for trouble-shooting purposes only.
Page 3: Sensor Errors
Sensor Error Hex Code
Menu item Format Description