86 Protec P3000 Messages
W41 HIGH FLOW too low For Protec P3000XL only in
HIGH FLOW mode:
Actual flow lower than lower
flow limit (for high flow mode) or
actual pressure at Wise
Technology sensor below
corresponding pressure limit
Filter in sniffer line blocked
Internal filter in main unit
blocked
Leak in
LOW FLOW capillary
Internal Leak
Replace sniffer tip filter (See
Section 7.4)
Replace internal filter*
Replace sniffer line!
Or: Replace probe cable!*
See Section 7
W42 Pressure at calibration
differs between
HIGH and
LOW FLOW
For Protec P3000XL only in
HIGH FLOW mode:
Pressure at Wise Technology
sensor differs from
LOW FLOW
mode to high flow mode (during
start-up or calibration only)
Filter in sniffer line blocked
Internal filter in main unit
blocked
Capillary blocked
Replace sniffer tip filter and
recalibrate (See Section 7.4)
Replace internal filter* and
recalibrate
Replace sniffer line and recalibrate!
Or: Replace probe cable!* and
recalibrate
See Section 7
W53 Flow at calibration out of
limits!
Only for Protec P3000 or
Protec P3000XL in LOW FLOW
mode:
Flow at calibration higher than
upper flow limit or lower than
lower flow limit or limit low flow
error (if enabled, whichever
comes first)
Filter in sniffer line blocked!
Internal filter in main unit
blocked
Leak in (low flow) capillary
Wrong flow limit or limit low flow
error settings (for
LOW FLOW
mode)
Replace sniffer tip filter and
recalibrate! (see section 7.4)
Replace internal filter* and
recalibrate!
Replace sniffer line (or: replace
probe cable*) and recalibrate!
Check flow limit settings and limit
low flow error (for
LOW FLOW mode)!
See section 4.4.1
Error
no.
Plain text message Possible Reason Recommendation