General Safety Precautions 9
Before connecting the Protec P3000 to the mains you must make sure that
the mains voltage rating of the Protec P3000 coincides with the locally
available mains voltage.
Do not suck in any liquids.
Permissible maximum input voltage PLC 28 V.
Permissible max. voltage and current for open collector outputs are: 28 V;
50 mA.
Maximum load rating relay outputs is 60 V DC / 25 V AC and 1 A per relay.
The air filter should be checked for contamination at least every 6 months
and should be definitely exchanged after 2 years.