tinb29e1-c (2019-10) VGC083A_B.om 81
10 Service
Every INFICON module is calibrated prior to shipment using nitrogen. Care should
be exercised when using gases other than nitrogen (N
2
) / air (see previous sections
regarding the use of gases other than N
2
/air).
Display on VGC083 controller indicates
No power Check cable connections and verify that
The connector may not be wired
Check cable connections and verify that
Readings appear very different from
expected pressure
Sensor not in the proper location on
vacuum system
The process gas is different from the gas
(Nitrogen) used to calibrate the IG
Ensure the sensor is located in
appropriate location on vacuum chamber
Apply gas correction factor if applicable
Sensor has been dropped causing
Replace the ion gauge sensor tube
The gauge sensor tube is contaminated
Degas the sensor or replace the sensor
Leak in the vacuum system Re-check for leak in the system. Re-
check that all metal seals are used when
operating below 1.00 × 10
-7
Ion gauge cannot be turned on Pressure exceeds 1.00 × 10
-3
Torr at
0.10 mA (100 µA) emission current
Decrease pressure below required value
Pressure exceeds 5.00 × 10
-4
Torr at
4 mA emission current
Pressure exceeds 1.00 × 10
-4
Torr at
Decrease pressure below required value
Decrease pressure below required value
Emission Control not functioning Switch to the other filament, replace ion
gauge sensor tube due to possible fila-
ment failure or contamination
Research Screen shows filament voltage
is present but filament current stays at
Filament is open Switch to the other filament, replace
sensor or electronics
Research Screen shows filament voltage
and filament current are present but
Filament is contaminated or burned out Switch to the other filament or replace
sensor
Unable to initiate degas System pressure above 5.00 × 10
Torr Decrease pressure below the required
Voltage to filament too high Filament contaminated or near end of its
Switch to the other filament or replace
Setpoint does not actuate
10.2 Troubleshooting - IG