70 tinb29e1-c (2019-10) VGC083A_B.om
When using the Log-Linear convection gauge analog output mode (Log-Linear
0 - 7 V, 1 V/decade) for gases other than N
2
/ Air, use the following look-up table
and information to convert the analog output to pressure. The look-up table has
been derived from the equation provided in section 7.8
True
Pressure
N
2
Ar He O
2
CO
2
KR Freon12 Freon22 D
2
Ne CH
4
Values listed under each gas type are in volts (V (dc)).
The gas in use is O
2
. Voltage output is 3.288 volts.
True pressure of O
2
is 2.00E-01 Torr
8.3.3.2 CG1 0 - 7 V or CG2 0 - 7 V
Analog Output Correction
Factors - Convection