EasyManua.ls Logo

Inficon VGC083A - 6.5.5 Analog Output Configuration

Inficon VGC083A
88 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
44 tinb29e1-c (2019-10) VGC083A_B.om
[Factory default = 0.0 SECONDS]
(REL 5 HI DELY: same info as REL 1 HI DELY above)
[Factory default = 0.0 SECONDS]
(REL 6 LO DELY: same info as REL 1 LO DELY above)
[Factory default = 0.0 SECONDS]
(REL 6 HI DELY: same info as REL 1 HI DELY above)
[Factory default = OFF]
This allows the user to manually toggle the relays on and off to test for correct
external circuit wiring and ensure polarity is as desired. Press ENTER to access
the TEST RELAYS menu. Select a specific relay by pressing the ENTER key so
that the relay number is highlighted. Press the UP or DOWN key to energize the
specific relay.
Relays will return to the previous state after exiting the TEST RELAYS menu.
The VGC083 provides three analog outputs proportional to the pressure measured
by the ionization gauge, the selected convection gauge (CG1 or CG2) or a continu-
ous, combined output signal from the IG and CG1 (IG - CG1) for full range meas-
urement. Press the ENTER key to access the ANALOG OUTPUT menu for config-
uring the analog outputs.
[Factory default = CG1 - CG1]
This menu provides various scaling and gauge type selections for Analog Output
#1. Use the UP and DOWN keys to select the desired analog output configuration
from the following available choices.
IG - CG1 0.5 - 7V
IG - CG2 0.5 - 7V
IG - ALT 0.5 - 7V
Log-linear analog output combining IG & CG1 as one signal
Log-linear analog output combining IG & CG2 as one signal
Log-linear analog output combining IG & Alternate gauge as
one signal
IG LOG N - 10
Log-linear analog output for IG only, output also compatible
with GP 307, 350
- IG LOG N - 11
Log-linear analog output for IG only, output also compatible
with GP 307, 350, 358
IG LOG N - 12
Log-linear analog output for IG only, output also compatible
with GP 307, 350
IG 1.8 - 8.7V
Log-Linear analog output for IG only
IG LINEAR
Linear analog output for IG only, output also compatible with
GP 307
ALT LINEAR
Linear analog output for alternate gauge only
ALT CG 1-8V
Log-linear analog output if alternate gauge connected is an
INFICON PGE300 or PGE500
ALT IG LOGN10
Log-linear analog output if alternate gauge connected is an
INFICON BAG302
CG1 1 - 8V
Log-linear analog output for CG1
CG2 1 - 8V
Log-linear analog output for CG2
CG1 0 - 7V
Log-linear analog output for CG1 , output also compatible
with GP 307,350,358
CG2 0 - 7V
Log-linear analog output for CG2 , output also compatible
with GP 307,350,358
CG1 NON - LIN
Non-linear analog output for CG1
CG2 NON - LIN
Non-linear analog output for CG2
CG1 LINEAR
Linear analog output for CG1
CG2 LINEAR
Linear analog output for CG2
REL 5 HI DELY
REL 6 LO DELY
REL 6 HI DELY
TEST RELAYS
6.5.5 Analog Output
ANALOG OUT 1

Table of Contents