42 tinb29e1-c (2019-10) VGC083A_B.om
Press the ENTER key to access the RELAYS menu for configuring the setpoint
relays.
[Factory default = ION GAUGE]
This assigns Relay #1 to the ion gauge, CG1, CG2 or ALT GAUGE (alternate
gauge).
[Factory default = CG1]
ASSIGN RELAY 2; same choices as ASSIGN RELAY 1 above
[Factory default = CG2]
ASSIGN RELAY 3; same choices as ASSIGN RELAY 1 above
[Factory default = ION GAUGE]
ASSIGN RELAY 4; same choices as ASSIGN RELAY 1 above
[Factory default = CG1]
ASSIGN RELAY 5; same choices as ASSIGN RELAY 1 above
[Factory default = CG2]
ASSIGN RELAY 6; same choices as ASSIGN RELAY 1 above
[Factory default = 1.00E-06]
This setpoint corresponds to the turn on point for Relay #1. Relay #1 will turn on
when the pressure drops below this setting. If you are unable to increase the value
of REL 1 LO TRIP (VGC083 responding with the message "NO CHANGE"), you
must first access the REL 1 HI TRIP menu below and increase that value to a
number higher than the value of the REL 1 LO TRIP you are trying to set. Ensure
the REL 1 LO TRIP setpoint value you are programming is within the measurement
range of the specific gauge the relay is assigned to.
[Factory default = 2.00E-06]
This setpoint corresponds to the turn off point for Relay #1. Relay #1 will turn off
when the pressure rises above this setting. If you are unable to decrease the value
of REL 1 HI TRIP (VGC083 responding with the message "NO CHANGE"), you
must first access the REL 1 LO TRIP menu above and decrease that value to a
number lower than the value of the REL 1 HI TRIP you are trying to set. Ensure the
REL 1 HI TRIP setpoint value you are programming is within the measurement
range of the specific gauge the relay is assigned to.
[Factory default = 1.00E-01]
(RELAY 2 LO TRIP; same info as RELAY 1 LO TRIP above)
[Factory default = 2.00E-01]
(RELAY 2 HI TRIP; same info as RELAY 1 HI TRIP above)
[Factory default = 1.00E-01]
(RELAY 3 LO TRIP; same info as RELAY 1 LO TRIP above)
[Factory default = 2.00E-01]
(RELAY 3 HI TRIP; same info as RELAY 1 HI TRIP above)
6.5.4 RELAYS
ASSIGN REL 1
ASSIGN REL 2
ASSIGN REL 3
ASSIGN REL 4
ASSIGN REL 5
ASSIGN REL 6
REL 1 LO TRIP
REL 1 HI TRIP
REL 2 LO TRIP
REL 2 HI TRIP
REL 3 LO TRIP
REL 3 HI TRIP