tinb29e1-c (2019-10) VGC083A_B.om 9
IG analog output
Log-linear
0 … 10 V (dc), 1 V/decade, or
1.79.3 V (dc) (nominal 1.88.7 V (dc)),
0.8 V/decade
Linear
0 … 10 V (dc) (useable over 3 decades)
Combination IG + CG analog
Log-linear
0.5 … 7 V (dc), 0.5 V/decade
CG analog output
Log-linear
1 … 8 V (dc), 1 V/decade, or
0 … 7 V (dc), 1 V/decade
Linear
0 … 10 V (dc) (useable over 3 decades)
Non-linear analog S-curve
0.3755.659 V (dc)
CDG one 0 … 10 V (dc) analog input signal from a
CDG when used as an alternate gauge to CG2,
or analog input from one of the following
INFICON modules: PGE300, PGE500,
BAG302
Protocol
RS485 / RS232 - ASCII protocol
IG sensor on/off status message is displayed on the front panel, by serial commu-
nications and by SPDT relay (DIGITAL I/O Connector) rated at 1 A at 30 V (dc)
resistive, or 1 A at 30 V (ac) non-inductive.
IG degas on/off status or IG error conditions are displayed on the front panel, by
serial communications and by an open collector transistor output (ground emitter)
rated at 40 V max. VCE, 50 mA IC max.
IG filament switching User selectable between filament 1 or 2 using
the front panel soft-keys or RS232 / RS485
serial comm
IG emission current
BAG050 / 051 / 052 / 053
100 μ
A, 4 mA, 10 mA automatic switching of
emission current (auto ranging)
BAG055 100 μA, 4 mA or automatic switching of
emission current (auto ranging)
IG degas
VGC083A with BAG050
VGC083A with BAG055
VGC083B with BAG051 /
052 / 053
40 W, electron bombardment
3 W, electron bombardment
40 W, resistive
IG overpressure protection
BAG050 / 051 / 052 / 053
Ion gauge turns off at the following default
settings:
1 × 10
-3
Torr at 100 μA emission current
5 × 10
-4
Torr at 4 mA emission current
1 × 10
-4
Torr at 10 mA emission current
BAG055 Mini IG 5 × 10
-2
Torr at 100 μA emission current
1 × 10
-3
Torr at 4 mA emission current
Analog output
Analog input
Serial communications
Status output
IG (Ionization Gauge)