EasyManua.ls Logo

Interacoustics Eclipse - 10 Technical specifications; 10.1 Technical specifications - Eclipse hardware

Interacoustics Eclipse
147 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
D-0120578-C – 2019/11
Eclipse Instructions for Use - US
Page 125
10 Technical specifications
10.1 Technical specifications - Eclipse hardware
Medical CE-mark:
The CE-mark indicates that Interacoustics A/S meets the requirements of Annex II of the
Medical Device Directive 93/42/EEC.
Approval of the quality system is made by TÜV identification no. 0123.
Standards:
Safety:
IEC 60601-1:2012, Class I, Type BF
CSA C22.2 No.60601-1:2014 and US ES 60601-1:2012.
EMC:
IEC 60601-1-2:2014
Power Supply:
Input Volts:
100 240VAC, 50/60Hz.
Consumption:
26W (0.3A Max)
Safety marking
ECM60UT31-XE0410 and cURus
Operating environment:
Operating Temperature:
15 35 °C (59 - 95°F)
Rel. Humidity:
30 90%
Ambient Pressure:
98kPa 104kPa
Transport & Storage:
Storage Temperature:
Transport Temperature:
Rel. Humidity:
0°C – 50°C (32°F - 50°F)
-20 50 °C (-4°F - 122°F)
10 95% (non condensing
Warm up time:
10 minutes at room temperature (20 °C) (68°F).
General
PC control:
USB:
USB 1.1 or 2.0 for input/output for computer communication.
Eclipse if fully operated from a PC
Construction:
Metal cabinet
Eclipse Dimensions
(L x W x H) 28 x 32 x 5.5 cm (11 x 12.6 x 2.2 Inches)
Eclipse Weight
2.5kg / 5.5 lbs excluding accessories

Table of Contents

Other manuals for Interacoustics Eclipse