To ensure compliance with the EMC requirements as specified in IEC 60601-1-2, it is essential to use
only the following accessories:
ITEM
MANUFACTURER
MODEL
EPA Preamplifier
Interacoustics
-
EPA3 Cable Collector
Interacoustics
-
EPA4 Cable Collector
Interacoustics
-
LBK 15 Loop Back Box
Interacoustics
LBK15
IP30 insert earphones
Interacoustics
Insert earphone
Shielded Headphone
Interacoustics
Headphone
Bone Conductor
Radio Ear
Bone
OAE Probe
Interacoustics
-
Cochlear Trigger Cables
Interacoustics
-
Conformance to the EMC requirements as specified in IEC 60601-1-2 is ensured if the cable types
and cable lengths are as specified below:
Description
Length
Screened
Mains Cable
2.0m
Unscreened
USB Cable
2.0m
Screened
EPA Preamplifier
2.5m
Screened
EPA3 Cable Collector
0.5m
Screened
EPA4 Cable Collector
50mm/0.5m/2.9m
Screened
LBK 15 Loop Back Box
2.0m
Screened
Insert earphones
2.9m
Screened
Shielded Headphone
2.9m
Screened
Bone Conductor
2.0m
Screened
OAE Probe
2.9m
Screened
Cochlear Trigger Cables
1.5m/5m
Screened

Table of Contents

Other manuals for Interacoustics Eclipse

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Interacoustics Eclipse and is the answer not in the manual?

Interacoustics Eclipse Specifications

General IconGeneral
CategoryMedical Equipment
ManufacturerInteracoustics
OAEYes
ABRYes
StimulusClicks, Tone Bursts, Chirps
TransducerInsert earphones, Headphones
ConnectivityUSB
ApplicationsNewborn hearing screening
SoftwareEclipse Software Suite
Battery LifeNot applicable (mains powered)