Guidance and manufacturer’s declaration — electromagnetic immunity
The OtoReadTM is intended for use in the electromagnetic environment specified below. The customer
or the user of the OtoReadTM should assure that it is used in such an environment,
IEC / EN 60601
test level
Electromagnetic environment
– guidance
Conducted RF
IEC / EN 61000-4-6
Radiated RF
IEC / EN 61000-4-3
3 Vrms
150kHz to 80 MHz
3 V/m
80 MHz to 2,5 GHz
3 Vrms
3 V/m
communications
equipment should be
used no closer to any
parts of the OtoReadTM,
including cables, than the
recommended separation
distance calculated from
the equation applicable to
the frequency of the
transmitter.
Recommended
separation distance
80 MHz to
800 MHz
800 MHz to
2,5 GHz
Where P is the maximum
output power rating of the
transmitter in watts (W)
according to the
transmitter manufacturer
and d is the
recommended separation
distance in meters (m).
Field strengths from fixed
RF transmitters, as deter-
mined by an
electromagnetic site
survey, (a) should be less
than the compliance level
in each frequency range
(b)
Interference may occur in
the vicinity of equipment
marked with the following
symbol:
NOTE1 At 80 MHz and 800 MHz, the higher frequency range applies
NOTE 2 These guidelines may not apply in all situations. Electromagnetic propagation is affected by
absorption and reflection from structures, objects and people.
(a)
Field strengths from fixed transmitters, such as base stations for radio (cellular/cordless) telephones
and land mobile radios, amateur radio, AM and FM radio broadcast and TV broadcast cannot be