OtoRead - Instructions for Use - US Page 65
predicted theoretically with accuracy. To assess the electromagnetic environment due to fixed RF
transmitters, an electromagnetic site survey should be considered. If the measured field strength in the
location in which the OtoReadTM is used exceeds the applicable RF compliance level above, the
OtoReadTM should be observed to verify normal operation, If abnormal performance is observed,
additional measures may be necessary, such as reorienting or relocating the OtoReadTM.
(b)
Over the frequency range 150 kHz to 80 MHz, field strengths should be less than 3 V/m.
Conformance to the EMC requirements as specified in IEC 60601-1-2 is ensured if the cable types
and cable lengths are as specified below:
Description
Length
Screened
OAE cable
2.0 m
Screened
USB Cable
2.0 m
Screened