7.5 Troubleshooting on the ACS Control System
Wheelchair slows down or doesn't move
Check: Possible cause: Direct remedy: Further action (s):
Status display on
joystick box
extinguished?
Power supply
interrupted.
Check plug
connection at
joystick box.
Contact authorised
dealer.
Battery fully
discharged.
Charge battery. See operating
instructions of
charger.
Battery defective. Replace battery. Contact authorised
dealer.
Status display on
joystick box flashing?
Various causes. Check error code. See Chapter 7.5.1
Red LED in battery charge display and status display flashing, driving mode display
showing a horizontal bar.
Check: Possible cause: Direct remedy: Further action (s):
Battery
discharged.
Charge battery. See operating
instructions of
charger.