7.5.1 Error Codes
The ACS electronics are capable to clear some faults automatically. In this case, the flashing of the
status display will extinguish. To achieve this, switch the joystick box off and on again several
times. Wait each time for about 5 seconds before switching the joystick box on again. If that does
not eliminate the fault, identify the error by means of the flash codes listed below.
Flash codes of status display on joystick box:
Flash code:
Meaning:
Direct remedy:
Further action (s):
1 x flash
Module defective. -
Contact authorised
dealer.
2 x flashes
Error in accessories. -
Contact authorised
dealer.
Lifter is elevated.
Lower lifter
completely.
-
3 x flashes
Right motor error.
Connection
loose/defective.
Check plug-and-
socket
connections.
Contact authorised
dealer.
4 x flashes
Left motor error.
Connection
loose/defective.
Check plug-and-
socket
connections.
Contact authorised
dealer.

Table of Contents