EasyManua.ls Logo

Invacare MK6i - Page 50

Invacare MK6i
127 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
50
Flash code:
Meaning:
Direct remedy:
Further action (s):
5 x flashes
Error/brake error in
right motor.
Connection
loose/defective or
motor defective.
Check plug-and-
socket
connections.
Contact authorised
dealer.
Motors disengaged.
Engage motors.
Switch joystick box
off and on again.
-
6 x flashes
Error/brake error in left
motor. Connection
loose/defective or
motor defective.
Check plug-and-
socket
connections.
Contact authorised
dealer.
7 x flashes
Battery fully
discharged.
Re-charge battery. Contact authorised
dealer.
8 x flashes
Battery potential too
high.
-
Contact authorised
dealer.
9 or 10 x flashes
Error in data transfer
between modules.
-
Contact authorised
dealer.
11 x flashes
Motors overloaded.
Switch joystick box
off and on again.
-
12 x flashes
Compatibility problems
between modules.
-
Contact authorised
dealer.

Table of Contents