System-Related Functions
WHour: the power statistics information, when the battery test mode
or battery simulation mode is turned on, the data will also be cleared.
Reset Meas: The zero-clearing operation, when the instrument is re-
powered or the source /load is switched it will be also performed.
Reset ReE: The zero-clearing regenerative power information
operation.
d. Press [Esc] to exit.
The information will display on the home screen when the Ext-Meter Disp
set to On in system menu, the operation refers to 7.4.7 Observe the DUT
Temperature and Total Historical Regenerative Power.
12.000V 10.000A
38.0 1000.0Wh
7.6 UUT Over-Temperature Protection
UUT over-temperature protection function allows the user to enable the
protection and set a over-temperature limit (Level). This function monitors the
temperature of UUT and prevents the UUT damaged due to higher
temperature. The operation procedures refer to 7.5 外部温度测量
An UUT over-temperature protection will occur if:
enable UUT OTP
The temperature of UUT is greater than the set protection point level.
Enable UUT OTP
1. Press the composite keys [Shift]+[R-set](Protect) to enter the Protect
menu.
2. Rotate the knob to select UUT OTP and press [Enter] to confirm.
3. Rotate the knob to select On and press [Enter] to confirm.
4. Rotate the knob and adjust the protection point Level and press [Enter] to
confirm.
5. Press [Esc] to exit.
Clear UUT OTP
When UUT OTP protection occurs, the instrument responds as follows:
Instrument output is off;
The buzzer sounds intermittently;
The interface displays “UUT Over Temperature” and flashes;
Status register, the UUT OT status bit is set to 1