EasyManua.ls Logo

ITech IT-M3900D Series - Page 153

ITech IT-M3900D Series
199 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Technical Specification
Parameter IT-M3902D-300-20
Read Back Temper-
ature Coefficient
Voltage ≤30ppm/
Current
≤50ppm/
Rise Time( no load ) Voltage
≤30ms
Rise Time( full load ) Voltage
≤60ms
Fall Time( no load ) Voltage
≤1s
Fall Time( full load ) Voltage
≤100ms
Dynamic response
time
(change from 20% to
90% of rated
current)
Voltage
≤1ms
Line Regulation
Voltage
≤0.01% + 0.01%FS
Current ≤0.03% + 0.03%FS
Load Regulation
Voltage
≤0.01% + 0.01%FS
Current ≤0.05% + 0.05%FS
Output protection
OCP 21A
OVP 303V
Overpower
protection
2040W
Remote Sense Com-
pensation Voltage
≤3V
External analog
optional
current program
External programming voltage 0V to 10V corre-
sponds to current 0A to 20A
current monitor
Current 0A to 20A corresponds to external moni-
toring voltage 0V to 10V
voltage program
External programming voltage 0 to 300V Corre-
sponding voltage 0 to 10V
voltage monitor
Voltage 0 to 300V corresponds to external moni-
toring voltage 0 to 10V
Other parameters:
AC Input
*2
Grid voltage range
3phase 200V480V
Single phase 100V240V
Frequency
50/60Hz
Max. apparent power
2.25kVA
Max. input current
12.5Aac
Copyright © Itech Electronic Co., Ltd.