Basic Operation
Category Parameter Initial Value
Upper limit of power: P+, and
lower limit of power: P-
Rated power value of
the instrument
[On/Off] status Off
Config
menu
Mode CV
Speed High
V-Rise Time/I-Rise Time 0.1s
V-Fall Time/I-Fall Time
Output Res 1000
On Delay/Off Delay 0
Last: Indicates when powered on, the instrument will remain the same
parameter settings (i.e., the parameters controlled by the *SAV 0 com-
mand, see 3.4 Save and Recall Operations) as last time you turned off
the instrument.
Last+Off: Indicates when powered on, the instrument will remain the
same settings as last time you turned off the instrument, but the output
state is Off.
Under the condition of Last+Off, the output switch state is not con-
strained by the *SAV 0 instruction.
4. After the parameter settings are complete, press [Enter].
For example, Last is selected, and the voltage value is set to 20V. After the
instrument is powered off and then powered on, the voltage displayed on the
interface is 20V.
3.6.3 Sense Function (Sense)
This menu item is used to switch the power supply to local measurement or re-
mote sensing.
The IT6000C series power supply supports two connection methods: Local
measurement and Remote sensing. The remote sensing is used for maximizing
measurement accuracy. (Refer to 1.4 Connecting the Device Under Test (DUT))
The procedures to set the menu item are as follows.