SOURce Subsystem
Copyright © Itech Electronic Co., Ltd. 91
Query Syntax
[SOURce:]POWer:RANGe? [ MINimum | MAXimum | DEFault ]
Returned Parameters
<NR3>
[SOURce:]POWer:HIGH
[SOURce:]POWer:LOW
This command sets the voltage high and low limit determined when in constant
power mode.
Command Syntax
[SOURce:]POWer:HIGH <NRf+>
[SOURce:]POWer:LOW <NRf+>
Parameters
MINimum through MAX | MINimum | MAXimum | DEFault
Unit
V (volts)
*RST Value
MAXimum
MINimum
Examples
POW:HIGH 5
Query Syntax
[SOURce:]POWer:HIGH? [MINimum|MAXimum|DEFault ]
[SOURce:]POWer:LOW? [MINimum|MAXimum|DEFault ]
Returned Parameters
<NR3>
[SOURce:]POWer:PROTection
This command sets the soft power protection level. If the input power exceeds
the soft power protection level for the time specified by POW:PROT:DEL, the