Loading...
SOURce Subsystem
Copyright © Itech Electronic Co., Ltd. 76
Examples
CURR:PROT:STAT 1
Query Syntax
[SOURce:]CURRent:PROTection:STATe?
Returned Parameters
<NR3>
Related Commands
CURR:PROT
[SOURce:]CURRent:PROTection
This command sets the soft current protection level. If the input current exceeds
the soft current protection level for the time specified by CURR:PROT:DEL, the
input is turned off.
NOTE: Use CURR:PROT:DEL to prevent momentary current limit conditions
caused by programmed changes from tripping the overcurrent protection.
Command Syntax
[SOURce:]CURRent:PROTection[:LEVel] <NRf+>
Parameters
0 through MAX | MINimum | MAXimum | DEFault
Unit
A (amperes)
*RST Value
MAXimum
Examples
CURR:PROT 2
Query Syntax
[SOURce:]CURRent:PROTection[:LEVel]?
Returned Parameters
NR3