Technical Specifications
Copyright © ITECH Electronic Co., Ltd. 61
voltage
mode
Resolution
1mV 10mV 1mV 10mV
Accuracy
±(0.025%+0.05%FS)
Constant
current
mode
Range
0~60A 0~600A 0~72A 0~720A
Resolution
1mA 10mA 1mA 10mA
Accuracy
±(0.1 %+0.1%FS)
Constant
resistance
mode *1
Range
0.01Ω~10Ω 10Ω~7.5KΩ 0.005Ω~10Ω 10Ω~7.5KΩ
Resolution
16bit 16bit
Accuracy
0.01%+0.08S *2 0.01%+0.0008S 0.01%+0.08S *2 0.01%+0.0008S
Constant
power
Mode *3
Range
4500W 6KW
Resolution
100mW 100mW
Accuracy
0.2%+0.2%FS 0.2%+0.2%FS
Dynamic mode
CC mode
T1&T2
20uS~3600S /Res:1uS
Accuracy
5uS±100ppm
Ascending/descending
slope *4
0.001~0.25A/uS 0.01~2.5A/uS 0.001~0.25A/uS 0.01~2.5A/uS
Minimum rise time *5
≒200uS
Measuring range
Read-back
voltage
Range
0~18V 0~120V 0~18V 0~120V
Resolution
1 mV 10 mV 1 mV 10mV
Accuracy
±(0.025%+0.025%FS)
Read-back
current
Range
0~60A 0~600A 0~72A 0~720A
Resolution
1mA 10mA 1mA 10mA
Accuracy
±(0.05%+0.1%FS)
Read-back
power
Range
4500W 6KW
Resolution
100mW 100mW
Accuracy
±(0.2%+0.2%FS)
Protection range
Overpower protection
≒4550W ≒6050W
Overcurrent protection
≒66A ≒66A ≒79.2A ≒792A
Overvoltage protection
≒130V
Overtemperature
protection
≒85℃
Specification
Short circuit
Current (CC)
≒66A ≒660A ≒79.2A ≒792A
Voltage (CV)
0V
Resistance
(CR)
≒3mΩ ≒2.5mΩ
Input
terminal
impedance
300KΩ
W439*H266*D590(mm)