1-2
1.2 Specifications
Light source : Hg-Xe lamp(150W)
Measurment wavelength range : 220-420 nm
Measurement range: 1000 mdeg (300nm) in 12 steps
Spectral bandwidth : 20 nm
Wavelength accuracy : ±5 nm
Sensitivity (S/N=2) : 0.05 mdeg/cm (290 m),
0.1 mdeg/cm (240 nm)
0.1g (Camphor-10-Sulfonic acid, 290 nm)
0.01g (Pantoy Lactone, 240 nm)
Baseline drift : 0.1 mdeg/cm/h (at 290nm)
Gain: x1, x10
Response : FAST, STD, SLOW
Flow cell path length : 25 mm(1,5,10 mm optional)
Signal outputs : Simultaneous outputs of CD
(circular dichroism) and UV absorption
CD, UV recorder output = 10mV full scale
CD integrator output = 1V/deg
UV integrator output = 1V/AU and 1V/2AU
Spectral measurement : Simultaneous measurement of CD and UV
spectra
Spectral memory : 220~420 nm(both UV and CD),
Memory:10files
Power requirement : 100-240 VAC, 400VA
Dimeniton: 390(W)x 450(D)x250(H)mm
Weight : approx. 21kg
The specifications are subjected to change without notice.