Care instructions KaVo uniQa
2 Reprocessing steps in accordance with ISO 17664 | 2.2 General instructions for the reprocessing of KaVo treat-
ment centres
9 / 56
Manual disinfection
4 Perform wipe disinfection only, the use of spray disin-
fection and disinfection baths is not feasible.
4 Wipe all outer surfaces with a soft cloth and an ap-
proved disinfectant.
KaVo recommends the following products based on com-
patibility of the materials. The microbiological efficacy must
be confirmed by the disinfectant manufacturer.
Application range in accordance with the available
manufacturer's instructions for use and national
guidelines.
Note the material safety data sheets.
Unless noted otherwise, approved disinfectants can be
used for manual wipe disinfection of the entire KaVo treat-
ment centre.
Approved disinfectants:
CaviCide / CaviWipes (Metrex)
Mikrozid AF Liquid (Schülke & Mayr)
FD 322 (Dürr Dental)
Incidin Liquid (Ecolab)
Automatic disinfec-
tion
KaVo recommends washer disinfectors in accordance with
EN ISO 15883-1 operated with alkaline cleaning agents.
The validations were conducted with a Miele washer disin-
fector using the VARIO-TD programme, the neodisher
MediClean cleaning agent and the neodisher Z neutralisa-
tion agent.
4 For programme settings as well as cleansers and disin-
fectants to be used, please refer to the Instructions for
Use of the washer disinfector.
4 In order to prevent impairment of the KaVo medical
device, make sure that the inside and outside of the
device is dry after the end of the cycle. Then service
the product in accordance with the product-specific
care instructions.
Manual drying
4 Supply dry, clean and uncontaminated compressed air
according to ISO 7494-2 only.
4 Blow off the outside and inside with compressed air un-
til water drops are no longer visible.
Automatic drying Usually, automatic drying is part of the cleaning/disinfec-
tion cycle of the washer disinfector and should not be car-
ried out above 95 °C.
Maintenance
4 Discard damaged products.
Control and func-
tional test
4 Perform a visual inspection to check for damage and
wear and tear.
4 Test products for function regularly.
Packaging
4 Use standardised packaging material. Sealing should
not be under tension.
4 Use sterile goods packaging in order to ensure sterility.
Sterilisation All sterilisable products are temperature-resistant up to
max. 138 °C.
Select a suitable procedure (depending on the available
autoclave) from the following sterilisation processes:
Autoclave with a triple pre-vacuum: at least 3 minutes
at 134 °C -1 °C/ +4 °C (273 °F -1.6 °F/ +7.4 °F)