2 Reprocessing steps in accordance with ISO 17664 | 2.4 Preparation of the dentist element
15 / 56
4 Pull-off the swing arm in an upward direction.
4 Clean the swing arm and roller manually.
4 Subject the swing arm S table to wipe disinfection.
2.4.4 Reprocessing of the instrument hoses
4 Clean the hoses and couplings with a cloth and water.
4 Then subject them to wipe disinfection.
CAUTION
Damage caused by spray disinfection.
Damage to the instrument hoses.
4 Do not subject instrument hoses to spray disinfection.
4 Subject instrument hoses to wipe disinfection.
2.4.5 Servicing the turbine return air filter
NOTE
The liquid collector must be checked weekly.
4 If there is any liquid in the filter, drain the liquid and replace the O-rings of
the MULTIflex couplings.