Care instructions KaVo uniQa
2 Reprocessing steps in accordance with ISO 17664 | 2.12 Reprocessing KaVo Screen One / KaVo Screen HD
46 / 56
4 Remove the cover.
4 Replace the amalgam container according to the instructions for use of
the amalgam separator.
2.12 Reprocessing KaVo Screen One / KaVo Screen HD
CAUTION
Product damage due to improper disinfection.
Malfunctions.
4 Use the disinfectant according to the specifications of the manufacturer.
4 No spray disinfection, perform wipe disinfection only.
4 Do not immerse product or parts of the product in liquids.
4 Mop up any spilled cleanser or disinfectant immediately.
4 Clean the surfaces with a soft cloth and water, and wipe-disinfect them with
a KaVo-approved surface disinfectant.
2.13 Reprocessing of the Service table 1568 and tray
holders
CAUTION
Product damage due to improper disinfection.
Malfunctions.
4 Use the disinfectant according to the specifications of the manufacturer.
4 No spray disinfection, perform wipe disinfection only.
4 Do not immerse product or parts of the product in liquids.
4 Mop up any spilled cleanser or disinfectant immediately.
NOTE
The handle on the service table can be taken off to improve the cleaning
properties.
4 Push the buttons on the bottom of the handle, and pull the handle off in
the direction of the arrow.