Care instructions KaVo uniQa
1 User instructions
5 / 56
1 User instructions
User guide
NOTE
Read these instructions prior to first startup to avoid misuse and prevent
damage. Comply with the Instructions for Use of the treatment centre.
Please note also the corresponding hygiene video on the KaVo website. Please
use the QR code or the following link:
https://www.kavo.com/hygiene-easy-way
General marks and symbols
CE- mark (European Community). A product bearing this mark meets
the requirements of the applicable EC directives.
Action request
Important information for users and service technicians
See Chapter on User Instructions/Hazard Levels
Thermodisinfectable
Sterilisation parameters
Steriliser with Dynamic-Air-Removal (pre-vacuum) method:
3 minutes at 135
o
C (275
o
F)
Drying time: 16 min.
Steriliser with gravity displacement method:
10 minutes at 135
o
C (275
o
F)
Drying time: 30 min.
Adhere to the expiry date
Lot de production
Original language German
Service
KaVo Technical Service
If you have any questions or complaints, please contact the KaVo Technical
Service:
+49 (0) 7351 56-1000
service.einrichtungen@kavo.com or service.treatmentunits@kavo.com