Care instructions KaVo uniQa
2 Reprocessing steps in accordance with ISO 17664 | 2.14 Reprocessing of the water bottle
47 / 56
1
1
4 Clean the surfaces with a soft cloth and water, and wipe-disinfect them with
an approved surface disinfectant.
Norm tray and non-slip mat can be sterilised.
Exemplary depiction of a tray holder
2.14 Reprocessing of the water bottle
CAUTION
Damage caused by defective water bottle.
Malfunctions or failures from a leaky water bottle.
4 Do not wash the water bottle in a dish washer, and do not rinse it with wa-
ter hotter than 50 .
4 Do not use the water bottle if it is scratched, deformed or discoloured.
4 Note the maximum useful life of the water bottle on the label.
Water bottle complete: (Mat. no. 0.710.4151)
Removing and cleaning the water bottle
4 Slowly turn the water bottle counterclockwise to take it off.