EasyManua.ls Logo

KaVo uniQa - Water Bottle Reprocessing: Operational Details and Disinfection; Water Bottle Over-pressure Release Mechanism; Daily Water Change and OXYGENAL 6 Recommendation; Water Bottle Cleaning Temperature Limit

KaVo uniQa
56 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Care instructions KaVo uniQa
2 Reprocessing steps in accordance with ISO 17664 | 2.14 Reprocessing of the water bottle
48 / 56
ð This allows the over-pressure to be released slowly from the water
bottle.
NOTE
The water in the water bottle needs to be changed at least once daily. In or-
der to ensure the efficiency of the measures, KaVo urgently recommends re-
processing the water with KaVo OXYGENAL 6. Also refer to: Care instructions
of the treatment centre.
4 For cleaning, rinse the water bottle with water at a temperature of less than
50 °C.
Manual internal disinfection
For manual internal disinfection, KaVo recommends the BC-San 100 liquid
made by Alpro.
Use this product according to manufacturer's instructions. For more informa-
tion, please refer to: https://www.alpro-medical.de/263-0-Gebrauchsanweisun-
gen.html