Care instructions KaVo uniQa
2 Reprocessing steps in accordance with ISO 17664 | 2.2 General instructions for the reprocessing of KaVo treat-
ment centres
10 / 56
▪ Autoclave with gravity procedure: at least 10 minutes
at 134 °C -1 °C/ +4 °C (273 °F -1.6 °F/ +7.4 °F)
4 Use according to the manufacturer's Instructions for
Use.
Storage
4 Reprocessed products should be stored protected from
dust with minimum exposure to germs in a dry, dark
and cool place.
4 Comply with the expiry date of the sterilised items.
Cleaning agents/
disinfectants
To ensure the product safety and maintain the quality of
the product, it is essential to use the following products:
▪ OXYGENAL 6 Mat. no. 0.489.3451 for continuous and
intensive germ reduction of the patient water in com-
bination with DVGW water block and DVGW water
bottle.
▪ DEKASEPTOL Gel Mat. no. 1.000.7205 for cleaning
and disinfection of the suction system.
2.2.1 Overview of reprocessing
CAUTION
Incorrect reprocessing
Property damage
4 Note table contents. Only approved components should be thermally disin-
fected and sterilised in a machine.
Reprocessing items
Medical devices
Cleaning
manual
Cleaning
auto-
matic
Disin-
fection
manual
Disin-
fection
auto-
matic
Steril-
isation
auto-
matic
Surfaces
Equipment X X
Cushion X X
Handle, white X X
Handle, grey X X X X X
Light X X
Monitor X X
Dentist element
Handpiece sleeve of the
three-function or multi-
functional handpiece
X X X
Cannula of the three-func-
tion or multifunctional
handpiece
X X X X X
Tray holders X X
Holder pad/non-slip mat X X X X X
Handpiece holder X X
Swing arm S-table X X
Handpiece holder for sur-
gery assembly kit
X X X X X
Saline pump assembly kit X X
Intraoral camera X X
Instrument hoses X X