Care instructions KaVo uniQa
2 Reprocessing steps in accordance with ISO 17664 | 2.7 Reprocessing the three-function and multifunctional
handpiece
22 / 56
2.7.3 Manual reprocessing
NOTE
Adapters are needed for manual reprocessing. Order adapter separately. Ad-
apter manual 3F/MF (Mat. no. 3.005.7780)
CAUTION
Never reprocess this medical device in an ultrasonic cleaner.
Malfunction and material damage.
4 Reprocess in a washer disinfector or by hand only.
CAUTION
Damage caused by using a nozzle needle that is too long.
Damage to the internal hoses of the cannula.
4 Use a short nozzle needle (Mat. no. 0.410.0921) to clean the media exit
holes.
Manual external cleaning
4 Clean the key part, gripping sleeve and cannula manually.
4 Carefully wipe off contamination from the cannula tip with a soft Q-tip or
soft cloth and water.
4 Carefully remove contamination from the media exit ports using the short
nozzle needle.
Manual internal cleaning
Validated manual internal cleaning (removal of residual protein) can be done
with KaVo CLEANspray.
4 Cover the medical device with the KaVo Cleanpac bag, and place it on the
cleaning adapter manual 3F/MF syringe cannula (Mat. no. 3.005.7780).
4 Hold the can vertically.
4 Press the spray button three times for 2 seconds each time.
4 Remove the medical device from the spray attachment and let the cleanser
act for 1 minute.
Also refer to:
Instructions for use KaVo CLEANspray