Care instructions KaVo uniQa
2 Reprocessing steps in accordance with ISO 17664 | 2.7 Reprocessing the three-function and multifunctional
handpiece
23 / 56
Manual external disinfection
WARNING
Incomplete disinfection.
Infection hazard.
4 Only use disinfection procedures that are verified to be bactericidal, fungi-
cidal and virucidal.
4 If the disinfectants/disinfection procedures fail to meet the mandatory na-
tional requirements, perform a final sterilisation with the sterilisation para-
meters as described.
CAUTION
Never reprocess the medical device with chloride-containing products.
Malfunction and material damage.
4 Reprocess by hand only.
Consumables required:
▪ Cloths for wiping the medical device.
4 Spray the disinfectant on a cloth, then wipe down the medical device and al-
low the disinfectant to act according to the instructions of the disinfectant
manufacturer.
4 Follow the instructions for use of the disinfectant.
Manual internal disinfection
CAUTION
Never reprocess the medical device with chloride-containing products.
Malfunction and material damage.
4 Reprocess by hand only.
WARNING
Incomplete disinfection.
Infection hazard.
4 Only use disinfection procedures that are verified to be bactericidal, fungi-
cidal and virucidal.
4 If the disinfectants/disinfection procedures fail to meet the mandatory na-
tional requirements, perform a final sterilisation with the sterilisation para-
meters as described.
The efficacy of manual internal disinfection must be demonstrated by the man-
ufacturer of the disinfection agent. With KaVo products, only use disinfection
agents that have been released by KaVo with respect to the compatibility of
materials (e.g. WL-cid / made by ALPRO).
4 Cover the medical device with the KaVo Cleanpac bag, and place it on the
cleaning adapter manual 3F/MF syringe cannula (Mat. no. 3.005.7780).
4 Hold the can vertically.
4 Press the spray key for at least 3 seconds.
4 Remove the medical device from the spray attachment and let the disinfect-
ant act for 2 minutes.
4 Follow the instructions for use of the disinfectant.
Manual drying
Use KaVo DRYspray for subsequent drying of the air and water ducts.