Care instructions KaVo uniQa
2 Reprocessing steps in accordance with ISO 17664 | 2.3 Reprocessing of the surfaces of treatment centre / uphol-
stery / foot control
11 / 56
Reprocessing items
Medical devices
Cleaning
manual
Cleaning
auto-
matic
Disin-
fection
manual
Disin-
fection
auto-
matic
Steril-
isation
auto-
matic
Assistant element
Connection parts of the
suction hose
X X X X
External suction hoses X X
Patient element
Tumbler support X X
Tumbler filler X X X X
Glass spittoon bowl
(smoke grey)
X X X X
Porcelain spittoon bowl X X X X
Instrument holder of the
hygiene centre
X X X X
Lid of the hygiene centre X X
Accessories
CENTRO support system X X
2.3 Reprocessing of the surfaces of treatment centre /
upholstery / foot control
Position the chair for cleaning
4 Tap the "Cleaning position" button in the "Cleaning" tab to drive the chair
into the cleaning position.
Also refer to:
Instructions for use of the treatment centre
Locking the touch screen
4 Tap the "Lock screen" button on the dentist element or the KaVo Logo on
the assistant element to lock the screen for wiping with a disinfectant.
ð The following display is shown on the dentist element:
Cleaning and disinfecting the surfaces of treatment
centre / upholstery
CAUTION
Non-colour-fast textiles.
Non-colour-fast textiles can cause permanent discoloration.
4 Immediately remove any discolouration with water and subsequently per-
form a wipe disinfection.