Chapter 2. Specifications
6 Kaye LTR -90 Operating Manual
2.2 Specifications
The following section lists the specifications for this instrument. Accuracy specifications are applicable for a one-year
calibration interval.
2.2.1 Base Unit Specifications
Part Numbers and Descriptions
Temperature Range at 23°C
-95°C to 140°C (-139°F to 284°F)
Display Accuracy
±0.2°C Full Range
Stability
±0.015°C Full Range
Loading Effect
(with a 6.35 mm reference probe and three 6.35 mm probes)
±0.006°C Full Range
(versus display with 6.35 mm probes)
±0.25°C at -95°C
±0.10°C at 140°C
Operating Conditions
0°C to 35°C, 0% to 90%
RH (non-condensing) < 2000 m altitude
Environmental conditions for all specifications except temperature range
13°C to 33°C
Immersion (Well) Depth
160 mm (6.3 in)
Well Diameter
30 mm (1.18 in)
Sleeve Hole Diameter
X0395 LTR -90 Low Temperature Dry Well -95° to 140°C, (115Volts, 60Hz)
X0395CE LTR -90 Low Temperature Dry Well --95° to 140°C , (230Volts, 50Hz)
X0395JP LTR -90 Low Temperature Dry Well -95° to 140°C, (115Volts, 50/60Hz)
X0395AS LTR -90 Low Temperature Dry Well -95° to 140°C (230Volts, 50/60Hz)