EasyManua.ls Logo

Kaye LTR -90 - 2.2 Specifications

Kaye LTR -90
50 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Chapter 2. Specifications
6 Kaye LTR -90 Operating Manual
2.2 Specifications
The following section lists the specifications for this instrument. Accuracy specifications are applicable for a one-year
calibration interval.
2.2.1 Base Unit Specifications
Part Numbers and Descriptions
Temperature Range at 23°C
-95°C to 140°C (-139°F to 284°F)
Display Accuracy
±0.2°C Full Range
Stability
±0.015°C Full Range
Loading Effect
(with a 6.35 mm reference probe and three 6.35 mm probes)
±0.006°C Full Range
(versus display with 6.35 mm probes)
±0.25°C at -95°C
±0.10°C at 140°C
Operating Conditions
0°C to 35°C, 0% to 90%
RH (non-condensing) < 2000 m altitude
Environmental conditions for all specifications except temperature range
13°C to 33°C
Immersion (Well) Depth
160 mm (6.3 in)
Well Diameter
30 mm (1.18 in)
Sleeve Hole Diameter
X0395 LTR -90 Low Temperature Dry Well -95° to 140°C, (115Volts, 60Hz)
X0395CE LTR -90 Low Temperature Dry Well --95° to 140°C , (230Volts, 50Hz)
X0395JP LTR -90 Low Temperature Dry Well -95° to 140°C, (115Volts, 50/60Hz)
X0395AS LTR -90 Low Temperature Dry Well -95° to 140°C (230Volts, 50/60Hz)

Table of Contents