Chapter 3. Quick Start
10 Kaye LTR -90 Operating Manual
3.2 Placement (cont.)
CAUTION! For safe operation and maintenance of the Product:
Energize the LTR -90 for a 2-hour dry-out period before use, if the LTR -90 was:
In transport
In a humid or semi-humid storage environment
Not energized for more than 10 days
If the product is wet or has been in a wet environment, take necessary measures to remove
moisture prior to applying power.
Always operate the LTR -90 on a flat, level, stable surface.
Do not store the LTR -90 at temperatures above 50 °C. The product has a refrigeration system
and contains gasses under pressure.
Do not turn the Product upside down. The inserts will fall out.
To prevent damage to the cooling system, do not tilt the LTR -90 on its side or upside down while
the LTR -90 is operating.
3.3 Connect to Mains Power
Use the 2-meter (6-foot) mains power cord to connect the Product to a 120 V ac or 230 V ac outlet rated for at least 10
amps.
3.4 Turn On the Product
1. Push the “I” side of the power switch on the front panel of the Calibrator.
2. Monitor the Startup screen for errors while the product turns on. If an error shows, contact Amphenol Advanced
Sensors Customer Service.
3.5 Change Language
To change the display language:
1. Push .
2. Push .
3. Push .
4. Push .
5. Push or to highlight a language.
6. Push to set language.