Kaye LTR -90 Operating Manual 39
Chapter 7. Maintenance and Troubleshooting
Chapter 7. Maintenance and Troubleshooting
7.1 Introduction
This chapter supplies information on how to service the Calibrator.
7.2 Maintenance
7.2.1 Clean the Product
Clean the Product surfaces with a damp cloth and mild detergent. Make sure liquids do not go into the Product.
7.2.2 Moisture Removal
When the Calibrator is run at a temperature below 0°C for a long period of time, moisture can build up in the well. This
moisture needs to be removed from the well.
To remove the moisture from the well and the insert probe holes:
1. Remove the rubber well insulator.
2. Set a 100°C SETPOINT (see Chapter 6).
3. Cool the unit back to ambient.
4. Clean the insert (see “Clean the Insert”).
Note: Before you clean or decontaminate the Product with a procedure not recommended by Amphenol Advanced
Sensors, speak with a Amphenol Advanced Sensors Customer Service associate for more information.
7.2.3 Clean the Insert
The insert must be removed and cleaned regularly to prevent the buildup of hard-water deposits. Hard-water deposits
can make it difficult to remove the insert from the Calibrator. If the insert is stuck in the well, use the special
instructions in the “Insert Installation and Removal Procedure” in Chapter 5.
To clean the insert, remove the insert with the removal procedure in Chapter 5. After the insert is cool, wipe the insert
with a cloth.